Bending of the UniTip 320µm and EndoTip 200µm:
1. Please sterilize the bending tool prior to each use to keep the sterile
conditions of the single-use fiber tip, see chapter "Cleaning,
disinfection and sterilization"
2. Now you can insert the single-use fiber tip into the bending tool and
bend the fiber tip to the angle that you need for best handling.
Adjusting the position of the endo stopper of EndoTip 200µm
The single-use fiber tip for endodontic applications (EndoTip 200µm) is
provided with an already mounted endo stopper, which serves as a stop
collar for the length of the root canal. The endo stopper can be moved on
the optical fiber.
1. To guarantee the sterile conditions of the fiber tip, please use the
sterilized bending tool to adjust the position of the endo stopper.
2. Place the EndoTip 200µm in the bending tool so that the endo
stopper positions itself in the notch for the stopper at the bending tool.
3. Adjust the position of the endo stopper by moving the handpiece up
and down so that the end of the fiber tip has the needed distance to
the stopper (indicated by the mm scale on the bending tool).
WARNING
The optical fiber of the single-use tip may be damaged if it is seriously
bent. This may constitute a health hazard for patients, dentists and
dental assistants.
Only use the original bending tool provided by the manufacturer to bend
the fibers properly.