Operating Instructions
båÖäáëÜ
cê~å´~áë
35
Adjusting the fiber length with the fiber cutter
Generally, all types of sterile delivered single-use fiber tips have the
proper fiber length in order to start working immediately without the need
of adjusting the length prior to application. However, in some cases it will
be necessary to adjust the length of the fiber.
Please sterilize the fiber cutter prior to each use to keep the sterile
conditions of the single-use fiber tip, see chapter "Cleaning, disinfection
and sterilization".
1. Place the optical fiber of the single-use fiber tip in the fiber cutter at
the notched mark.
2. Press the fiber cutter together and release it again.
3. Bend the optical fiber at the notched location.
The optical fiber breaks at the notched location with a smooth,
perpendicular fracture surface.
Check to see if the light of the aiming beam projects a uniform circular
pattern. To do this, aim the optical fiber vertically at a white background.
If the probe projects no pattern at all or only an uneven pattern, cut off
another one to two millimeters.
After treatment
The easiest and safest way to dismount the UniTip 320µm or EndoTip
200µm from the handpiece after treatment is to use a disposable
container.
1. Open the lid of the disposable container and connect the plastic grip
of the tip to the suitable recess inside the container.
2. Remove the single-use fiber tip from the handpiece by pulling the
container away from the handpiece.
3. The single-use fiber tip falls into the disposable container.
4. Close the disposal container.
NOTICE
Press firmly but do not squeeze the optical fiber. You just need a little
notch to receive a perfect result when breaking the fiber at the notch.
WARNING
If the optical fiber of the single-use fiber tip does not protrude at least
10 mm out of its metal tube, there is a risk that the tube will heat up.
CAUTION
As soon as you disassemble the single-use fiber tip after treatment
make sure to protect the optical fiber socket with the protective cap
provided for this purpose. Make sure that no dust or dirt can enter the
optical system. Otherwise the unit may be permanently damaged.