Settings
Failure to use the settings specified in this manual or perform the actions
described here may lead to a dangerous exposure to radiation.
The manufacturer cannot be held liable for any damage caused by
improper use or non-compliance with the instructions and information
provided in this manual.
Flammable materials
Never direct the laser beam towards any flammable material, e.g. paper
or plastics. They could catch fire due to the high temperatures produced
by the laser beam.
The unit is not suitable for use in the presence of anesthetics that are
flammable when in contact with air, oxygen or nitrogen monoxide.
Oxygen-saturated materials such as cotton wool can catch fire due to
high temperature that the unit reaches during operation. Label removers
and flammable solutions used for cleaning and disinfecting the
Aquamarine should be allowed to evaporate before using the device.
Observe fire hazards caused by flammable gases.
Reducing the risk of burns
Unintentional exposure to the laser radiation may lead to burns. This risk
can be reduced by surrounding the target area with moistened sterile
drapes or moistened gauze. These materials must meet the requirements
of laser surgery.
2.6
Nominal ocular hazard distance
The nominal ocular hazard distance (NOHD) from the distal end of the
optical fiber is 1.5 m.
CAUTION
Use of controls or adjustments or performance of procedures other than
those specified herein may result in hazardous radiation exposure.