Operating Instructions
båÖäáëÜ
cê~å´~áë
13
2.8
UniTip, EndoTip and ActiveTip
Optical fiber tips and connection socket
Make sure that no dust, dirt and foreign particles can enter the optical
fiber socket or the optical system. Never place your finger or any other
objects in the optical connectors. Otherwise the unit may be permanently
damaged.
When disconnecting the disposable fiber tips (UniTip 320µm, EndoTip
200µm) or the glass rod (ActiveTip 8mm) from the Aquamarine, always
cover the connection socket at the handpiece with the special protection
cap supplied. Make sure that the optical system is clean before
connecting the disposable fiber tip or glass rod.
The optical fiber must not be twisted inside the tube of the single-use fiber
tip (UniTip 320µm and EndoTip 200µm). There is a risk of breakage.
Stop the laser activation of the Aquamarine immediately if the UniTip
320µm, EndoTip 200µm or ActiveTip is broken. Otherwise the tips may
become hot.
UniTip 320µm, EndoTip 200µm and ActiveTip must be checked for sure
seating prior to each use.
2.9
Contamination
Accessories
Danger of (cross) contamination. Pay attention not to hurt or stick yourself
or any other person with the laser fiber tip. This applies also if the
handpiece is placed in the holder.
Prior to each use, the handpiece sleeve and the optical light guide
(ActiveTip) must be sterilized. The single-use fiber tips are delivered
sterile and must be used only once.
Tissue particles
During cutting and coagulation of tissue, tissue particles could become
aerosolized. Always wear a face mask, because a risk of infection exists.
A extractor or a filter should be used. The operating personnel should be
aware that biologically active material could get into the environment. It
may contain particles of viable tissue.
WARNING
Single-use fiber tips (UniTip 320µm, EndoTip 200µm) must not be
sterilized again after usage. They are disposable products.
➢ Never reuse the single-use fiber tips as they can also demage the
handpiece optics if used more than one time.