Aquamarine
2.3
Operating personnel
Qualification/education
The Aquamarine may only be operated by educated and qualified
personnel (dentist, assistent, dental hygienist). The applicable
occupational safety regulations and accident prevention measures, the
current operating instructions and national requirements concerning
education must be complied with.
Know-how
Know-how and expertise about laser therapy as well as the skilled use of
the laser and the applied indications are required. Please refer to
applicable country-specific requirements.
Experience
The manufacturer recommends to gain practical experience in laser
dentistry before first use of the Aquamarine by attending an appropriate
training. Amongst others FONA Dental offers trainings. Please see the
FONA Dental homepage (www.fonadental.com).
Obligation of the user
Users are obliged to use only faultless materials, to ensure correct
application and to protect themselves, the patient and other persons
against hazards.
Unauthorized access
In order to prevent false or improper use, the Aquamarine must not be
used by unauthorized persons. Therefore the Aquamarine equipment
must be protected against unauthorized access when not in use. This can
be achieved, for example, by switching the Aquamarine off after usage so
that the electronic access key (pin code) must be entered before using it
again.
2.4
Physical working principle
The 445 nm laser radiation of the Aquamarine is generated via a laser
diode inside the handpiece and guided to the treatment region via quartz
fibers. The laser radiation is absorbed by the tissue and converted to heat
used for cutting, coagulation and germ reduction.
2.5
Laser radiation hazards
Never direct the laser or aiming beam toward a person's eye! All persons
present in the room e.g. patient, dentist and assistant must always wear
the laser protective goggles.
Observe all labels on the Aquamarine.
Master switch of the practice
Note that after switching off the master switch of the practice the
Aquamarine will still remain switched on. It is then energized by the
rechargeable battery.
Emergency stop
In case of an emergency press the "Laser Stop" button to the right of the
touch screen on the front side of the Aquamarine control unit. The breaker
always cuts off the voltage from the device.
WARNING
The Aquamarine may only be used and maintained by thoroughly
trained personnel.