Calibration
87
6.1.2
Powder components (MP and powder additive)
Calibrate the powder feed components as described in the pre-
vious chapter.
6.2
Semi and fully automatic calibration of the feeding stations with peristaltic pumps
6.2.1
Semi-automatic calibration
The semi-automatic calibration process of a station with peristal-
tic pumps is exemplified by
F-station 1
:
1.
>
calibration
>
F-station 1
2. Confirm
start?
with
. The semi-automatic calibration pro-
cess starts running and a series of automated processes
takes place.
3. After completion of the calibration process, the message on
the right will be displayed.
4. Now also calibrate the other stations which are equipped with
the peristaltic pumps.
Note
: If no precision scales is available for the calibra-
tion of powder additives, you have to repeat the calibra-
tion process several times in order to get a larger
additive amount. Divide the measured quantity by the
number of calibration processes and then key in the fig-
ure.
Repeat the calibration process in order to check the re-
sult.
Note
: The peristaltic pumps must be calibrated only if
they are used for
parallel feeding.
<F-station 1 >
start?
...
calibration
completed!
Содержание TAK5-CH1-25
Страница 38: ...38 Operation ...
Страница 54: ...54 Start up ...
Страница 70: ...70 Setup ...
Страница 84: ...84 Device data ...
Страница 90: ...90 Calibration ...
Страница 150: ...150 Cleaning ...
Страница 170: ...170 Animal control ...
Страница 192: ...192 Failures and warnings ...
Страница 196: ...196 Care and maintenance plan Routine tasks ...
Страница 215: ......
Страница 218: ...218 Annex 16 3 5 Template for individual feeding plan ...
Страница 219: ...Annex 219 16 4 Basic principle of interval feeding ...
Страница 222: ...222 Annex ...