Operating Manual EMGZ 321.EIP
19.10.2017
36
1)
The display shows the unit of measure that was previously selected
3)
Permanent key pressing expedites the changing speed of the value.
4)
This parameter is only displayed, if the function check “Limit 1” or “Limit 2” have
been assigned to Relay 1 respectively Relay 2.
5)
A relay function can only be assigned once to a relay output. If a function is being
used the function will not appear on the parameter selection.
8.7
Description of Relay Parameters
Relay Output 1
Relay Output 1
LCD:
Relay 1
LCD:
Relay 2
Use:
Each relay output can individually be configured to check violations of
force limits (see also 3.10 “Relay Outputs”). If a violation is detected
an event can be initiated (e.g. activate a flash light or alarm horn). A
relay function can only be assigned once to a relay output. If a function
is being used the function will not appear on the parameter selection.
Unit
Parameter Range
Selection
Default
Min
Max
- -
-
Chk A
Chk B
Chk A+B
Chk IA-BI
Disabled
Relay Output 1=
Chk A+B
Relay Output 2=
Chk
׀
A-B
׀
Chk A
Check and indicate if force limit of sensor A has been violated.
Chk B
Check and indicate if force limit of sensor B has been violated.
Chk A+B
Check and indicate if the sum of the forces of sensors (A+B) has
been violated.
Chk
׀
A-B
׀
Check and indicate if the difference of the forces of sensors (
׀
A-B
׀
)
has been violated.
Disabled
Disable the relay output.