T+/T+PRO English Instruction Sheet
Page 8
Removing the Probe Tips
To remove the probe tips:
1.
Grasp the probe firmly in one hand.
2.
Hold the test lead jack firmly in the other hand.
3.
Pull to disengage the test lead probe from the test lead jack.
LED Display Voltages
Table 4 shows the differences in the displayed voltages on the
LEDs on the Canadian and USA versions of the Tester.
Table 4. LED Display Voltages
USA Canada
Q
Q
R
R
S
S
T
T
U
U
V
V
W
X
X
Y
Y
Self Test
XW
Warning
To avoid possible electric shock or personal injury:
•
Perform a self test before any measurements are
taken for voltage, continuity, resistance, or rotary
field.
•
When the batteries are depleted, self test will not
function. Replace the batteries.
W
Caution
Although the Tester may be used with depleted batteries,
replace depleted batteries immediately to avoid Tester
damage from battery acid leakage.
To perform the self test, short the two probes. The self test lights
all LEDs, and all LCD segments (T+ PRO). The beeper sounds
(unless disabled) and the Tester switches to continuity mode. The
self test will then be de-activated for 30 seconds. If the self test
needs to be repeated, wait 30 seconds.