Transcutaneous Pacemaker Testing
Setup
and
Testing
3
3-31
10.
Press
F3 REVIEW UP
or
DOWN
arrow key to view deviations falling
outside the selected deviation values for rate and amplitude.
Up to 200 deviations will be stored in the Analyzer memory. Any long
term test exceeding 200 deviations is automatically terminated. The pulse
width at the “out of limit” pacemaker pulse is also reported.
11.
Press
F4 XMIT
to transmit data.
12.
Press
F1 RESTART
to restart the test.
13.
Press
ESC
to return to previous menu(s).