SigmaPace 1000
Operators Manual
3-8
P ac ing R a te
Inter
v
al
R es et > 750 mS
P R P
S R P
P ac ing R a te
Inter
v
al
R es et (2 X )
P R P
"S ens ed" C ardia c
T es t P
u
ls e #1
S R P
"S ens ed" C ardia c
T es t P
u
ls e #1
"S ens ed" C ardia c
T es t P
u
ls e #2
eyr004.eps
Figure 3-3. Sensed Refractory Period
Long Term Test
The overall stability of the pacemaker’s pulse rate and amplitude are
monitored continuously over an extended time period. Each measured pulse is
compared to factory preset limits of 10 % for both monitored variables. Test
limits are adjustable by the user.