SigmaPace 1000
Operators Manual
3-28
flexibility allows the user to enter either the actual test pacemaker values or the
selected rate and amplitude settings.
Deviation values are independently selectable from 2 % to 20 % in 1 %
increments. The Analyzer compares these incoming pacer pulses to the target
values. When the long-term test is complete, output pulses from the
pacemaker that were not within the selected limits may be reviewed.
F1
F2
F3
F4
eyr013.eps
1.
From the
SELECT TEST
menu, press
F2 LONG TERM
.
F1
F2
F3
F4
eyr026.eps
2.
Press
F2 RATE
to toggle between pulse rate and amplitude. This
parameter works with the
LIMIT
parameter to assign limits to both.