SigmaPace 1000
Operators Manual
3-26
F1
F2
F3
F4
eyr066.eps
8.
Press
F4
XMIT DATA
to transmit data.
9.
Press
F1
RELEASE
to release the hold and return to the
HOLD
screen.
10.
Press
ESC
to return to previous menu(s).
Purpose of the ECG Simulation Test
This test lets the technician check external pacemakers for basic operation,
performance, and intermittent problems. Additionally, it is an excellent
training tool for clinical staff in the setup and use of external pacemakers.
F1
F2
F3
F4
eyr013.eps
1.
From the
SELECT TEST
menu, press
F1
ECG SIM
.