SigmaPace 1000
Operators Manual
3-20
4.
Press
F1 RESTART
to restart the test.
5.
Press
F4
XMIT DATA
to transmit data.
6.
Press
ESC
to return to previous menu(s).
Amplitude of ECG Signal for Demand Mode Pacemaker
The following test determines the amplitude of the simulated ECG signal
required by the demand mode pacemaker.
F1
F2
F3
F4
eyr012.eps
1.
From the
SELECT TEST
menu, press
F2
SENSE AMP
.
F1
F2
F3
F4
eyr062.eps