Introduction and Specifications
Instrument
Specifications
1
1-25
Resolution
............................................................. ±1 LSD
Display Format
...................................................... 3 digits
Physiological Simulation
Selection (Default) .............................................. Single Channel Simulation (V)
Square Wave
Ventricular Channel ............................................ Simulated R-wave
Width................................................................... 1.0 mS
Amplitude............................................................ 20 mV
peak
Active Outputs .................................................... Ventricular Channel
(4 mm Banana Jacks) Only
Available Test Load(s) ........................................ 500
Ω
Only
Additional Waveform Selections
Square (SQU)
Triangle (TRI)
Haversine (HSN)
Sine Square (SSQ)
Asymmetrical Triangle (ISO) (Fixed Width2 mS rise time / 13 mS fall time)
Amplitude
Pacer Load Selection.......................................... 500
Ω
Range ............................................................. 0.05 mV
peak
(50 V
peak
) to
50.0 mV
peak
Accuracy......................................................... ±5 % of setting
Resolution (Step Size).................................... 0.05 mV
peak
steps from 0.05 mV
peak
to
0.95 mV
peak
0.50 mV
peak
steps from 1.0 mV
peak
to
50.0 mV
peak
Default Setting ................................................ 20.0 mV
peak
Width
Range ................................................................. 0.15 mS to 300 mS
Accuracy ............................................................. 5 % of setting
Selection Count .................................................. 58
Resolution (Step Size)........................................ 0.05 mS steps from 0.15 mS to
0.95 mS
1.0 mS steps from 1.0 mS to 19.0 mS
5.0 mS steps from 20 mS to 95.0 mS
25 mS steps from 100 mS to 300 mS
Default Setting (mS) ........................................... 30 mS
Intended Pacemaker Types................................ VVI (Ventricular Pace and Sense
Only)
Compatible Pacemaker Rates ............................ 20 to 200 PPM
Содержание SigmaPace 1000
Страница 6: ......
Страница 10: ...SigmaPace 1000 Operators Manual iv...
Страница 12: ...SigmaPace 1000 Operators Manual vi...
Страница 14: ...SigmaPace 1000 Operators Manual viii...
Страница 16: ...SigmaPace 1000 Operators Manual 1 2...
Страница 46: ...SigmaPace 1000 Operators Manual 2 2...
Страница 56: ...SigmaPace 1000 Operators Manual 3 2...
Страница 86: ...SigmaPace 1000 Operators Manual 3 32...
Страница 88: ...SigmaPace 1000 Operators Manual 4 2...
Страница 128: ...SigmaPace 1000 Operators Manual 4 42 F1 F2 F3 F4 eyr052 eps 10 Press F1 END to end the test F1 F2 F3 F4 eyr053 eps...
Страница 130: ...SigmaPace 1000 Operators Manual 4 44...
Страница 132: ...SigmaPace 1000 Operators Manual 5 2...
Страница 154: ...SigmaPace 1000 Operators Manual 5 24...