Transvenous Pacemaker Testing
Setup
and
Testing
4
4-23
F1
F2
F3
F4
eyr042.eps
10.
Press
F4
XMIT DATA
to transmit data.
11.
Press
F1
RESTART
to restart the test.
12.
Press
ESC
to return to previous menu(s).
Amplitude of ECG Signal for a Demand Mode
Pacemaker
This test determines the required amplitude of the simulated ECG signal to be
detected by the demand mode pacemaker.
F1
F2
F3
F4
eyr033.eps
1.
From the
SELECT TEST
menu, press
F2
SENSE AMP
.