SigmaPace 1000
Operators Manual
1-22
Available Selection(s)......................................(3) 200, 500 and 1000
Ω
±1 %
Default Setting.................................................500
Ω
Amplitude
Pacer Load Selection ..........................................500
Ω
Range..............................................................0.05 mV
peak
(50 V
peak
) to 50.0 mV
peak
Accuracy..........................................................±5 % of setting
Resolution .......................................................0.05 mV
peak
steps from 0.05 mV
peak
to
0.95 mV
peak
0.50 mV
peak
steps from 1.0 mV
peak
to
50.0 mV
peak
Pacer Load Selection ..........................................200
Ω
Range..............................................................0.05 mV
peak
(50 V
peak
) to 20.0 mV
peak
Accuracy..........................................................±5 % of setting
Resolution .......................................................0.05 mV
peak
steps from 0.05 mV
peak
to
0.95 mV
peak
0.50 mV
peak
steps from 1.0 mV
peak
to
20.0 mV
peak
Pacer Load Selection ..........................................1000
Ω
Range..............................................................0.05 mV
peak
(50 V
peak
) to 100 mV
peak-to-
peak
Accuracy..........................................................±5 % of setting
Resolution .......................................................0.05 mV
peak
steps from 0.05 mV
peak
to
0. 95 mV
peak
0.50 mV
peak
steps from 1.0 mV
peak
to
49.5 mV
peak
05.0 mV
peak
steps from 50 mV
peak
to
100 mV
peak
Default Setting
.......................................................2.5 mV
peak
Width
Range..................................................................0.15 mS to 300 mS
Accuracy..............................................................±5 % of setting
Selection Count ...................................................58
Resolution (Step Size).........................................0.05 mS steps from 0.15 mS to 0.95
mS
1.0 mS steps from 1.0 mS to 19.0 mS
5.0 mS steps from 20 mS to 95.0 mS
25 mS steps from 100 mS to 300 mS
Intended Pacemaker Type(s)
VVI (Atrial Pace and Sense Only)
Compatible Pulse Rates......................................30 to 200 PPM
Содержание SigmaPace 1000
Страница 6: ......
Страница 10: ...SigmaPace 1000 Operators Manual iv...
Страница 12: ...SigmaPace 1000 Operators Manual vi...
Страница 14: ...SigmaPace 1000 Operators Manual viii...
Страница 16: ...SigmaPace 1000 Operators Manual 1 2...
Страница 46: ...SigmaPace 1000 Operators Manual 2 2...
Страница 56: ...SigmaPace 1000 Operators Manual 3 2...
Страница 86: ...SigmaPace 1000 Operators Manual 3 32...
Страница 88: ...SigmaPace 1000 Operators Manual 4 2...
Страница 128: ...SigmaPace 1000 Operators Manual 4 42 F1 F2 F3 F4 eyr052 eps 10 Press F1 END to end the test F1 F2 F3 F4 eyr053 eps...
Страница 130: ...SigmaPace 1000 Operators Manual 4 44...
Страница 132: ...SigmaPace 1000 Operators Manual 5 2...
Страница 154: ...SigmaPace 1000 Operators Manual 5 24...