Page 3.65
=============================================================
RECALL SETUP RS
-------------------------------------------------------------
Purpose:
Recalls an internally stored setup. This setup must have
been stored in the ScopeMeter manually or with the SS
(Save Setup) command.
The effect of the RS command is that the instrument setup
is recalled and the instrument forced to running state.
Command Syntax:
RS <setup_reg><cr>
where,
<setup_reg> = 1 to 15 : Screen/Setup memories
1001 : Long Record/Replay memory Input A
1002 : Long Record/Replay memory Input B
Response Syntax:
<acknowledge><cr>
Note: The new setup is active when you have received the
<acknowledge> response from the ScopeMeter.
Example:
The following example program saves the present setup in
setup memory 8. You are requested to change the present
settings. Then the original settings are recalled from
setup memory 8 and made the actual setting.