Page 3.28
=============================================================
PROGRAM COMMUNICATION PC
-------------------------------------------------------------
Purpose:
Programs the baud rate for RS232 communication:
Command Syntax:
PC <baudrate>
where,
<baudrate> = 1200|2400|4800|9600|19200
38400 (Fluke 19xC)
57600 (Fluke 19xC, PM9080/101 required)
The default baudrate is 1200. This is set at power-on or
after a Reset Instrument command (command "RI")
Notes:
The Fluke 19x/19xC instruments support 1 stopbit,
8 databits and software handshake (X-on X-off
protocol).
Hardware handshaking is not supported.