Page A.2
Synchronization Error
Returned when the ScopeMeter receives data while it
does not expect any data. This can occur as follows:
- The ScopeMeter receives a new command while a
previous command or query is not yet completely executed.
You can prevent this error by doing the following:
1. Read the <acknowledge> reply after each command or
query.
2. If this <acknowledge> is zero and if a query was sent to
the ScopeMeter, read all available response
data.
Communication Error
Any framing, parity or overrun error detected on the received
data will cause Communication Error.