Rotating non Ex Rated Enclosures
The gland which couples the sensing probe to the enclosure provides a critical sealing function
for the enclosure. Internal wires are passed through this gland and the high integrity seal. This
gland incorporates a Union Joint which is designed to rotate.
However, this rotation is limited to one-time adjustment of Display orientation after installation on
site.
ONLY one 36mm spanner applied to the Hex of Union Joint to rotate enclosure to desired orien
-
tation as allowed. DO NOT hold the enclosure during this procedure.
Rotation beyond these strict limits can damage the internal cables
Ensure Enclosure follows the spanner rotation and assembly integrity is not compromised
Содержание TRACER 1000 3G LTT1-3G
Страница 15: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 15 ...
Страница 18: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 18 ...
Страница 20: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 20 ...
Страница 21: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 21 ...
Страница 22: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 22 ...
Страница 23: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 23 ...
Страница 24: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 24 ...
Страница 25: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 25 ...