Adjusting Probe Length
Rigid Probes
Cut rigid probes to appropriate length. After adjustment, you must change the 'ProbeLength' Parameter in 'Advanced'
menu to represent the new length (password 222).
Flexible Probes
(a) Mark the point at which the flexible cable enters the cable weight.
(b) Release the cable weight grub screws with hex key.
(c) Measure and note the length of cable concealed within cable weight.
(d) Cut cable noting the length of cable must include the concealed length above.
(e) Re-insert the cable into the weight and tighten grub screws to tightening Torque of 20Nm. Use loctite 243 or
equivalent on grub screws to secure once completed.
(f) Adjust ProbeCalibr Parameter in 'Advanced' menu to represent new length (password 222).
Rotating Ex Rated Enclosures
The Ex d gland which couples the sensing probe to the flameproof enclosure provides a critical sealing
function for the enclosure. Internal wires are passed through this gland and the high integrity seal. This
gland
incorporates a Union Joint which is designed to rotate.
However, this rotation is limited to one-time adjustment of Display orientation after installation on site.
ONLY one 36mm spanner applied to the Hex of Union Joint to rotate enclosure to desired orientation as
allowed. DO NOT hold the enclosure during this procedure.
Rotation beyond these strict limits can damage the internal cables
Ensure Enclosure follows the spanner rotation and assembly integrity is not compromised
The gland which couples the sensing probe to the enclosure provides a critical sealing function
for the enclosure. Internal wires are passed through this gland and the high integrity seal. This
gland incorporates a Union Joint which is designed to rotate.
However, this rotation is limited to one-time adjustment of Display orientation after installation on
site.
ONLY one 36mm spanner applied to the Hex of Union Joint to rotate enclosure to desired orien
-
tation as allowed. DO NOT hold the enclosure during this procedure.
Rotation beyond these strict limits can damage the internal cables
Ensure Enclosure follows the spanner rotation and assembly integrity is not compromised
Содержание TRACER 1000 3G LTT1-3G
Страница 15: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 15 ...
Страница 18: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 18 ...
Страница 20: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 20 ...
Страница 21: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 21 ...
Страница 22: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 22 ...
Страница 23: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 23 ...
Страница 24: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 24 ...
Страница 25: ... 877 356 5463 p 330 331 7331 f 330 331 7172 www FLO CORP com 2018 FLO CORP REVA 0918 25 ...