4.4. Automated Mode
FLIR Detection
Countdown
Figure 97.
Countdown announcing the start of a scheduled measurement
Automatic identi ication measurements override other activities except alarms (see
, p.
). Au-
tomatic measurements are blocked during:
• Background measurements in Finder Mode (see
, p.
• A running self test (see
, p.
During these conditions, the countdown continues, but “Blocked” is shown instead of the count-
down timer. Should the condition end before the scheduled time, the measurement starts. Other-
wise, the measurement is omitted and a new attempt is made after the regular interval.
During the measurement, basic radiation data and a small spectrum with logarithmic scaling are
shown (Figure
98
, p.
94
).
Progress Bar
Figure 98.
Automated measurement