FLIR Detection
6. FLIR identiFINDER R300 Menu Reference
Figure 176.
Dose rate alarm threshold settings
Warning
The value with unit for the warning threshold. Use “Off” (the lowest value) to suppress this
alarm. Available units are
nSv/h – µSv/h – mSv/h – Sv/h (nrem/h – µrem/h – mrem/h – rem/h)
𝜎
Alarm
The sensitivity of the detection of unusual gamma radiation increases (see
5.3
, p.
101
).
You can specify “Off” to suppress the noti ications or a value from 1.1 to 9.9. The lower the
value, the better the sensitivity for radiation increases but the more noti ications you can
expect. Higher values are recommended to reduce false noti ications in environments with
generally low radiation levels.
Down
,
Select
Choose and select menu items, see
2.2.2
, p.
35
for details.
+
,
-
,
Accept
,
Cancel
Edit values, see
2.2.3
, p.
37
for details.
Left
,
Right
Edit composite values, see
2.2.4
, p.
40
for details.
Exit
Return to the superior menu.
6.35
Neutron Alarm Thresholds
This item is available for FLIR identiFINDER R300 variants with a neutron detector (see
,
p.
identiFINDER
®
R300/en/2014.4(13623)/Feb2015
163