A.7. SPP Commands and Updates
FLIR Detection
Dose units (Lines 12 and 13)
DoseWarningThres
and
DoseAlarmThres
use dose units. The possible scaled rem units
are
”rem”
,
”mrem”
,
”µrem”
, and
”nrem”
. The possible scaled Sievert units are
”Sv”
,
”mSv”
,
”µSv”
, and
”nSv”
.
Dose rate units (Lines 6, 10 and 11):
DoseRate
,
DoseRateWarningThres
, and
DoseRateAlarmThres
use dose rate units. The
possible scaled rem per hour units are
”rem/h”
,
”mrem/h”
,
”µrem/h”
, and
”nrem/h”
. The possible scaled Sieverts per hour units are
”Sv/h”
,
”mSv/h”
,
”µSv/h”
, and
”nSv/h”
.
Operating Mode (Line 9):
OperatingMode
indicates the current mode the FLIR identiFINDER R300 is operating in. Pos-
sible values include:
”DoseRate”
,
”Finder”
,
”Identify”
, or
”Other”
.
Neutron Information (Lines 14, 15, 26, 27, and 29):
NeutronRateWarningThres
,
NeutronRateAlarmThres
,
NeutronCountRateWarning
,
Neu-
tronCountRateAlarm
, and
SigmaNeutronAlarm
are only available on instruments that con-
tain an He-3 tube.
Sigma Alarms (Lines 28 and 29):
Sigma alarms are currently available only on R300 products.
A.7.3 Get Spectra
The ”GetSpectra” information can be queried through the command speci ied below or is automat-
ically transmitted upon selection of ”send” on the user interface of the device.
Listing A.8.
GetSpectra Command
1
< ?
xml v e r s i o n
=
” 1.0 ”
e n c o d i n g =
” utf -8 ”
? >
2
< C o m m a n d n a m e =
” G e t S p e c t r a ”
>
3
< A r g u m e n t n a m e =
” N u m b e r ”
v a l u e =
”
(SpectrumNumber)
”
/ >
4
< / C o m m a n d >
\0
Line3: Spectrum Number
If the GetSpectra command has an
Argument
element then it must contain a
name
attribute
with a value of
”Number”
and a
value
attribute with a numerical value specifying the record
number of the spectrum to send.
Omit the
Argument
element to retrieve highest spectrum number.
Listing A.9.
GetSpectra Update
1
< ?
xml v e r s i o n
=
” 1.0 ”
e n c o d i n g =
” utf -8 ”
? >
252
identiFINDER
®
R300/en/2014.4(13623)/Feb2015