Specifications
SYMBOLS and SPECIFICATIONS
Operator's Manual
|
2-8
Patient Range
Adult - Pediatric (i.e. infant, child & adolescent) with body
weight >10 kg
Factory Default Parameters
Patient Settings
MODE
A/CMV
Volume Control
500 mL
t
I
1.0 sec
ƒ
15 b/min
Ptrig
-1.0 cmH
2
O
Paw Alarms
5 cmH
2
O Paw
40 cmH
2
O Paw
I
Alarms
3 L/min
I
20 L/min
I
PEEP/CPAP
0 cmH
2
O
Psupport
0 cmH
2
O
Humidifier
Off
Buzzer Volume
Loud
User Set Up
Power Save
Pressure Units
Auto Panel Lock
Set Up
On
cmH
2
O
Disabled
User
Patient Circuit
Reusable 22 mm I.D. adult/pediatric circuit with 3/16 inch
(4.8 mm) I.D. proximal pressure sensing line, 1/8 inch
(3.2 mm) I.D. exhalation valve control drive line, and
exhalation valve.
Exhalation Valve
FLIGHT
MEDICAL’S
F
LIGHT
50 exhalation valve (P/N
V54-00000-
67) is manufactured and designed specifically for the FLIGHT
50
Ventilator. FLIGHT MEDICAL does not approve of the use of
any type or brand of exhalation valve that has not been tested
and approved by FLIGHT MEDICAL for use with the
FLIGHT
50.
FLIGHT 50 HUMIDIFIER
(operates on A.C. power only)
Set Target Temperature Range
19ºC to 39ºC
Operating Water Volume
300 mL
Usable Volume of Water Bottle
265 mL
Compliance at Minimum Water Level
0.5 mL/cmH
2
O / mbar @ 23ºC