Operator's Manual
|
3-32
Pneumatic Requirements
: Oxygen 35-90 psig (2.4 to 6.2 Bar)
3.7.2 Oxygen Blending Bag Kit
The Oxygen Blending Bag Kit is used to blend atmospheric air with a low
flow (0 to 10 L/min) medical grade oxygen source. The Oxygen Blending
Bag Kit attaches to the Fresh Gas Intake on the Filter Cover, located on the
right side of the ventilator. This system allows the user to ventilate patients
with oxygen enriched gas from 21% to 100% FiO2.
Pneumatic Requirements:
Oxygen 0-10 L/min
3.7.3 Auto Lighter Cable
Figure 45 – Auto Lighter Cable
The Auto Lighter Cable (p/n A01-00040-29) allows the FLIGHT 50 Ventilator
to be powered through the D.C. lighter plug in an automobile. The internal
battery charges whenever the FLIGHT 50 Ventilator is connected to an
external power source, equal to or greater than 12 VDC, including the D.C.
lighter plug.
3.8
User Set Up
The User Set Up allows the operator to select a variety of functional
parameters. User Set Up parameters must be established before the FLIGHT
50 Ventilator is used for ventilation.
HOME CAREGIVERS: The User Set Up parameters should be
defined in conjunction with your physician or Homecare
Dealer. Once established, these parameters are kept in
memory and will be in affect each time the ventilator is
powered on until the operator enters User Set Up and makes
changes.
To enter User Set Up the FLIGHT 50 Ventilator must be in
the Standby condition.
To access User Set Up, when connected to A.C. power:
1. Press and hold the
Silence/Reset
button down.
2. When operating on internal battery, you must press the
On/Standby
button while pressing and holding the
Silence/Reset
button.
3. When the Message Display Window briefly displays the message
“USER
SETUP”,
release the button.