Operator's Manual
|
3-17
The value in the adjacent display window blinks.
2. Use the
▲Up/▼Down
controls to adjust the displayed alarm setting
value.
3. The new setting can be retained by:
Pressing the selected button again to accept the alarm setting
Selecting another parameter for adjustment
Allowing five seconds to elapse without adjustment.
In SPONT mode the Paw alarm is inactive but the value can
be preset.
The Paw alarm setting cannot be a value below PEEP/CPAP +
3. The Paw alarm setting must be a value at least 1 above the
Paw alarm setting.
3.3.2
▲ Paw (High Pressure) Alarm (user adjustable)
Figure 24 –
▲
Paw (High Pressure) Alarm
Audible Alarm:
Intermittent beep
Visual Alarm:
▲
Paw indicator blinks red
Message Window:
HIGH PRESSURE
The High
▲
Paw Alarm is activated when airway pressure (Paw) reaches the
Paw alarm limit setting. Any breath in progress immediately cycles to
exhalation.
The alarm violation is cancelled when Paw falls below the Paw alarm limit
setting and at least one second has elapsed since the alarm was activated.
3.3.3
▼ Paw (Low Pressure) Alarm (user adjustable)
Figure 25 –
▼
Paw (Low Pressure) Alarm
Audible Alarm:
Intermittent beep
Visual Alarm:
▼
Paw indicator blinks red
Message Window:
LOW PRESSURE