Safety Instructions
Cautions
Operator's Manual
|
1-3
Only a FLIGHT MEDICAL approved exhalation valve can be
used with the FLIGHT 50 Ventilator.
Perform an exhalation valve calibration each time a clean
circuit/exhalation valve is installed.
The FLIGHT 50 Ventilator has been tested and found to comply with the
EMC limits according to the EN60601-1-1-2 standard class B. These limits
are designed to provide reasonable protection against harmful interference
in a typical medical installation. The equipment generates uses and can
radiate radio frequency energy and, if not installed and used in accordance
with these instructions, may cause harmful interference to other devices in
the vicinity. However, there is no guarantee that interference will not occur
in a particular installation. If it is found that this equipment causes harmful
interference with other devices, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the distance between the equipment.
Connect the equipment into an outlet on a circuit different from
that to which the device (s) is connected.
Consult the manufacturer for help.
1.2
Cautions
Only use medical grade oxygen with the Air/Oxygen
Entrainment Mixer or Oxygen Blending Bag Kit.
Do not place liquid containers in the immediate vicinity or
on top of the FLIGHT 50 Ventilator. Liquids that get into
the ventilator can cause equipment malfunction and
damage.
After the FLIGHT 5050 Ventilator is serviced, it must
completely pass an Operational Verification Procedure
(OVP) before being returned to patient use.
An authorized FLIGHT MEDICAL factory-trained technician
must do all service or repairs performed on the FLIGHT
5050 Ventilator.