9 Measurement
FLUXUS F808
9.3 Start of the measurement
97
UMFLUXUS_F808V2-2EN, 2020-06-25
9.2.5
Input of the error delay
The error delay is the time after which an error value will be sent to an output if no valid measured values are available.
This display will only be indicated if the list item
edit
is selected in
Special Funct.\Dialogs/Menus\Error-val.
delay
.
If the error delay is not entered, the damping factor will be used.
• Enter a value for the error delay.
• Press ENTER.
9.3
Start of the measurement
• Select the program branch
Measuring
.
• Press ENTER.
If the parameters in the program branch
Parameter
are
not valid or incomplete, the error message
NO DATA!
will
be displayed.
If the data logger or the serial interface is activated, the
measuring point number has to be entered:
Input of the measuring point number
• Enter the number of the measuring point.
• Press ENTER.
For the activation of text input, see
Special Funct.\SYSTEM settings\Dialogs/Menus\Meas.Point No.
Input of the sound path number
A value for the number of sound paths corresponding to the connected transducers and the entered parameters is recom
-
mended.
• Change the value, if necessary.
• Press ENTER.
Output Options\...\I1:Error-val. delay
Measuring
Meas.Point No.
Sound Path
Transd. Distance
Signal test
Transd. Distance
Measurement display
Measuring\...\Meas.Point No.
Measuring\...\Sound Path