6 Installation
6.2 Transducers
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
38
Ratchet clasp
• Insert the tension strap into the tension strap clamp (2) and the metal spring, see Fig. 6.21 and Fig. 6.24. The metal
spring does not have to be mounted on:
– steel pipes
– pipes with an outer pipe diameter < 80 mm
– pipes that are not subjected to significant temperature fluctuations
• Position the ratchet clasp, tension strap clamp (2) and metal spring (if necessary) on the pipe:
• Position the metal spring opposite to the tension strap clamp.
• Place the tension strap around the pipe and pass it through the slot of the clasp screw, see Fig. 6.25.
• Tighten the tension strap.
• Cut off the protruding tension strap, see Fig. 6.25.
• Tighten the clasp screw.
• Tighten the nut of the tension strap clamp (2), but not too firmly in order not to damage the tension strap, see Fig. 6.21.
Fig. 6.22: Tension strap with band
clamp clasp
Fig. 6.23: Tension strap with quick
release clasp and metal spring
1 – clasp screw
1 – clasp screw
2 – metal spring
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment. Observe the applicable rules.
Notice!
In order to release the screw and the tension strap, press the lever down, see Fig. 6.18.
Fig. 6.24: Tension strap with metal spring and clamp
1 – metal spring
2 – tension strap clamp
1
2
1
1
2