
9 Measurement
9.2 Measurement settings
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
96
• Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Zero-Scale Val.
is the value assigned to the value
Output MIN
of the output range.
• Press ENTER.
• Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
Full-Scale Val.
is the value assigned to the value
Output MAX
of the output range.
• Press ENTER.
Function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
yes
to test the output.
• Press ENTER.
• Enter a test value for the selected physical quantity. If the external measuring instrument displays the entered value, the
output functions correctly.
• Press ENTER.
• Select
yes
to repeat the test.
• Press ENTER.
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.
: 0 m³/h
Full-Scale Val.
: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA
Output Options\...\I1:Test output?
Output Options\...\I1:Test value
Output Options\...\I1:Test output?
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.
: 0 m³/h
Full-Scale Val.
: 300 m³/h
Test value
= 150 m³/h (center of the measuring range, corresponds to 12 mA)
If the multimeter displays 12 mA, the current output functions correctly.