16 Advanced functions
FLUXUS F808
16.10 Modification of the limit for the inner pipe diameter
135
UMFLUXUS_F808V2-2EN, 2020-06-25
16.10 Modification of the limit for the inner pipe diameter
It is possible to modify the lower limit of the inner pipe diameter for a given transducer type.
• Enter the HotCode
071001
, see section 8.4.
• Enter the lower limit of the inner pipe diameter of the displayed transducer type.
• Press ENTER to select the next transducer type.
16.11 Transducer temperature
It is possible to output the transducer temperature.
• Enter the HotCode
007043
.
• Select
yes
to display the transducer temperature during the measurement.
• Press ENTER.
• Select
yes
to store the transducer temperature.
• Press ENTER.
Notice!
If a transducer is used below its recommended inner pipe diameter, a measurement might be impossible.
DNmin Q-Sensor
15
mm
Show T-transd.?
no
>YES<
Store T-transd.?
no
>YES<