16 Advanced functions
16.13 Behavior of the alarm outputs
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
140
If the alarm state indication is activated, the alarm output state is displayed after configuring the alarm outputs.
The alarm state indication is structured as follows:
Rx =
, with
X
being the number of the alarm output and
a pictogram according to Tab. 16.3.
The configuration of the alarm outputs can be repeated by pressing key C. When the configuration of the alarm outputs is
finished, press ENTER. The main menu will be displayed.
If the alarm output indication is activated, the alarm state can be displayed during the measurement. Press key
to
scroll through the upper line or key
in the lower line until the alarm state is displayed.
16.13.5 Deactivation of an alarm output
If the programmed outputs are no longer required, they can be deactivated. The configuration of a deactivated output is
stored and will be available if the output is activated again.
• Select
no
in
Output Options\Alarm Output
to deactivate an output.
• Press ENTER.
Tab. 16.3: Pictograms for the alarm state indication
no.
func
(switching condition)
typ
(holding behavior)
mode
(switching function)
current state
R
=
OFF
NON-HOLD
NO Cont.
closed
1
MAX
HOLD
NC Cont.
open
2
MIN
3
+→- -→+
QUANT.
ERROR
Example
R1 =
Output Options\...\Alarm Output