background image

8

F. 

Replacing the Incidental (Top) Illuminator LED bulb on Middle/High School Stereo Models 
S71005A and S71005C:

a.  To replace top light replace top light, remove two chrome crosshead screws securing top 

lighthouse to light bracket and the two chrome crosshead screws securing coiled cord 
clamp to arm.   

b.  Remove front lens from lighthouse by rotating in a counter clockwise direction.   Holding 

top lighthouse in one hand, remove silver spacer then feed coiled cord through bottom of 
lighthouse, pushing LED lamp assembly out of lighthouse.

c. 

Unplug connector located on bottom of LED lamp mount and remove from housing.

d. 

Using Fisher Science Education replacement lamp no S74022 (LED bulb with special 
base) plug the connector to supply cord. Wipe bulb to insure that it is clean and free of 
fingerprints.

e.  Reassemble lamp assembly in reverse order

G.  Replacing the Transmitted (Bottom) Illuminator LED bulb on Middle/High School Stereo Models 

S71005A and S71005C:

a. 

To replace transmitted (bottom light), loosen setscrew located at side of base and remove 

stage plate.   Removing stage plate will expose the bottom LED bulb.  

b. 

Grasp LED bulb and pull straight up and out of socket. 

c. 

Using Fisher Science Education replacement lamp no S74023 (LED bulb with special 
base) insert the new LED bulb making sure to align the lamp base with lamp socket.  Wipe 
bulb to insure that it is clean and free of fingerprints. 

d.  Replace the stage plate and tighten locking setscrew.

H. 

Replacing Halogen Lamps on College/Laboratory Models S71003, S71003A, S71003B, 
S71003C, S71003D, S71003E, S71003F, S71003G, S71003H, S71003J, S71012 and S71012A:

a.  Carefully lay instrument on its side, taking care to avoid damage to specimen slide holder 

located on top of mechanical stage.

b.  Loosen large chrome screw located on hinged door of illuminator base.
c.  Swing door open to expose the halogen lamp.
d. 

Using a tissue or cloth, gently grasp the halogen bulb and pull lamp straight out of lamp 

socket.  

e. 

Using Fisher Science Education replacement lamp no S74020 hold lamp with tissue paper 

and insert bulb base pins straight into lamp socket. Wipe bulb to insure that it is clean and 

free of fingerprints. 

f. 

Make sure to use the proper 12-volt bulb in order to prevent serious damage to the 

electrical system of the microscope.

g.  Close hinged door and tighten the chrome locking screw

I. 

Replacing the Transmitted (Bottom) Lamp on High School/College Stereo Models S71007 and 
S71007A:

a. 

Gently lay microscope on its side to reveal base plate on bottom of illuminator base. 

a. 

Observe screws located in rubber feet. Using a screwdriver, remove the rubber feet and 

base plate to expose bulb. 

b.  Remove bulb by depressing lamp slightly and rotating in a counter-clockwise direction 

until it pops up.

c. 

Using Fisher Science Education replacement lamp no S74015 (120volt 15watt double 
contact base) insert new lamp by gently pressing into socket and rotate in a clockwise 

direction until it snaps into place.  Wipe bulb to insure that it is clean and free of all 

fingerprints.

d.  Replace base plate and secure rubber feet.

J.  Replacing the Incidental (Top) Lamp on High School/College Stereo Models S71007 and 

S71007A:

a.  Remove top lamp housing by grasping it and rotating in a counter-clockwise direction to 

expose the lamp.

Содержание S7001B

Страница 1: ...Microscope Instruction Manual...

Страница 2: ...nses The image produced by the compound microscope is upside down and reversed Compounds are used for viewing standard 1 by 3 1mm thick transparent specimen slides with cover slips Prepared specimen s...

Страница 3: ...C Instructions 21 Model S71003D S71003E S71003F S71003G S71003H and S71003J Instructions 22 General Instructions for Stereo Microscopes 23 25 Model S71004 and S71004A Instructions 26 Model S71005 S710...

Страница 4: ...Disc located below stage with holes of various apertures designed to help achieve optimum resolution of the objective lens Larger apertures used for higher magnifications and smaller apertures used f...

Страница 5: ...st to the object being viewed forms first image of the specimen OIL IMMERSION LENS High power 100x objective lens which requires a medium of immersion oil between the lens and the slide RESOLVING POWE...

Страница 6: ...cleaning will be submitted for cleaning III MECHANICAL MAINTENANCE A Rack Stop compound microscopes only Rack stop screw has been pre adjusted at the factory and should not require re adjustment Howev...

Страница 7: ...and replace locking screw C Replacing Incandescent Lamps on Middle High School Models S71001 and S71001A a Make sure that the illuminator field lens housing and lamp are cool before attempting to remo...

Страница 8: ...3B S71003C S71003D S71003E S71003F S71003G S71003H S71003J S71012 and S71012A a Carefully lay instrument on its side taking care to avoid damage to specimen slide holder located on top of mechanical s...

Страница 9: ...l it slips into position d Replace stage plate and secure with locking setscrew M Replacing the Incidental Top Illuminator Halogen Bulb on Digital Stereo Model S71012B a Remove front lens housing from...

Страница 10: ...il slot engages depress fuse cap and rotate turn to lock into fuse holder S Recharging Batteries on Cordless Models a Must use the supplied Automatic Switching Recharger when charging batteries b It i...

Страница 11: ...Locate battery compartment cover on bottom of illuminator Carefully remove the two screws securing battery cover to bottom of illuminator Slide cover back to expose batteries d Remove all 3 batteries...

Страница 12: ...Incorrect bulb voltage or lamp base used Replace with specified lamp Light flickers Lamp not properly inserted into socket Loose connection at AC outlet Electrical short Properly insert lamp Have qual...

Страница 13: ...slide dirty Spots on field of view Eyepiece or condenser lens dirty No immersion oil used on 100X objective lens Bubbles air in immersion oils Adjust disc diaphragm or iris diaphragm Adjust light int...

Страница 14: ...entered over lens in center condenser lens located in middle of stage B Mechanical Stage Models Swing movable finger on holder outward place specimen slide cover slip facing up on top of slide against...

Страница 15: ...ything when looking through the microscope A slight adjustment of focus and diaphragm openings will be needed to sharpen the image when changing objective magnification A Models supplied with 100x oil...

Страница 16: ...hat you will want adequate ambient light in order to illuminate your opaque specimen from the top If the specimen is small and falls through the aperture at center of stage place a small piece of pape...

Страница 17: ...ble on model S71000D Battery replacement for models S71000D See Replacing Batteries located on page 10 U MODEL S71000E and S71000F These two models have a special inclined eyepiece tube so that it is...

Страница 18: ...1D Replacing batteries see page 10 T Fuse replacement on models S71001 and S71000A See Replacing Fuses located on page 10 R MODEL S71001E S71001F AND S71001G Dual viewing head permits simultaneous vie...

Страница 19: ...ated on page 10 T S71002A AND S71002B STUDENT MODEL These models feature coarse and fine focusing mechanisms full sized DIN 4X 10X and 40X objectives widefield 10X eyepiece and a 0 65 condenser with i...

Страница 20: ...Fuse replacement on models S71002C and S71002E See Replacing Fuses located on pages 10 R S71002G HIGH SCHOOL AND COLLEGE This model features a Seidentopf binocular head coaxial coarse and fine focusin...

Страница 21: ...e To operate this series of microscopes see General Instructions for use of a Compound Microscope located on pages 14 and 15 Lamp replacement For Halogen lamp replacement see page 8 H Fuse replacement...

Страница 22: ...of objective lenses After properly focusing instrument adjust ocular tube diopters to compensate for the difference between left and right eyes Lamp replacement For Halogen lamp replacement see page 8...

Страница 23: ...ion Select illumination that meets the requirements of specimen and turn ON illuminator A There are three rocker type light controls located on microscope base a MAIN Turns power ON and OFF b I Turns...

Страница 24: ...e for this difference and assure that you will see one corrected image when looking through microscope a Observe that the knurled diopter ring on the left eyepiece tube can be rotated to move the eyep...

Страница 25: ...knob to the lowest magnification by aligning the number 1 located on the objective turret for model S71010 or by aligning the number 1 located on zoom control knobs to the index dot on the viewing hea...

Страница 26: ...bottom lamp located on page 9 L Fuse replacement See Replacing Fuses located on pages 10 R MODEL S71004A Features an adjustable post mounted viewing head with inclined eyepiece tubes for ease in viewi...

Страница 27: ...harge batteries prior to using this instrument See Recharging Batteries page 10 Q To use and focus this microscope see General Instructions for Stereo Microscopes located on pages 23 25 Models S71005...

Страница 28: ...age 9 K Transmitted bottom lamp replacement See Replacing the transmitted bottom lamp located on page 9 L Fuse replacement See Replacing Fuses located on page 10 R MODELS 71007 and S71007A Features in...

Страница 29: ...Fuses located on page 10 R MODEL S71008A Fixed arm with inclined viewing head Illuminated by a variable halogen incidental top and a fluorescent transmitted bottom lighting system Change magnificatio...

Страница 30: ...located on page 9 O Fuse replacement See Replacing Fuses located on page 10 R MODEL S71010 Fixed arm with inclined viewing head Illuminated by a variable halogen incidental top and a fluorescent trans...

Страница 31: ...tic Images software allowing the user to view capture label manipulate and measure images In order to achieve optimum results it is important that you carefully read the instructions for software loca...

Страница 32: ...outlet and converter automatically switches to proper output voltage Refer to Quick Start Guide located on your CD for installation of the software and operation of the camera The three position slid...

Страница 33: ...matically switches to proper output voltage Refer to Quick Start Guide located on your CD for Installation of the software and operation of the camera The three position sliding rod located on the sid...

Страница 34: ...plug into 12VDC switching power converter then insert plug on other end of converter into power jack on back of microscope base Plug AC cord into either a 120v or 240v outlet and converter automatical...

Страница 35: ...ble magnification from10X to 40X Illumination Insert power plug into 12VDC switching power converter then insert plug on other end of converter into power jack on back of microscope base Plug AC cord...

Страница 36: ...cluded with microscope Instructions on connecting to your wireless enabled laptop or desktop are covered in the Wifi Camera Operation section of this manual MODEL S71014 This WiFi digital microscope t...

Страница 37: ...images on iOS or Android devices with the free MotiConnect App or through IP address using web browser and the camera s on board software You must charge batteries prior to using this instrument See...

Страница 38: ...warranty b Once the connection has been established you can begin using the WiFi camera through the MotiConnect App automatic or through your web browser using the following IP address http 192 168 1...

Страница 39: ...nician s discretion Other than set forth above Fisher Science Education hereby disclaims all warranties expressed or implied of fitness for a particular purpose Defective products covered by the warra...

Страница 40: ...tomer service call 1 800 955 1177 To fax an order use 1 800 955 0740 To order online www fisheredu com 2013 Thermo Fisher Scientific Inc All rights reserved Trademarks used are owned as indicated at w...

Отзывы: