GENERAL USE OF WIFI MICROSCOPES

In preparation for using the S71014, S71014A and S71014B, you may wish to download and install the 
free MotiConnect app. The MotiConnect app can be downloaded from the Google Play Store or Apple 
App Store, depending on whether you are using an Android or iOS device.   

  

Instructions for the MotiConnect app can be found within the application through the help button.  
Further instruction is available online from Motic and National Optical YouTube pages.  

 
Your WiFi camera also has built-in software, which is accessible by typing the cameras IP address 
into your web browser. These instructions are covered in the WiFi Camera Operation section of this 
manual.  
 

If you would like to connect the WiFi camera to a wireless enabled laptop or computer, install Motic 

Images software included with microscope. Instructions on connecting to your wireless enabled laptop 

or desktop are covered in the Wifi Camera Operation section of this manual.  

MODEL S71014

This WiFi digital microscope transmits up to 1.3MP live 
images to iOS or Android devices and features built-in 
high-speed, full resolution imaging technology. Use it as 

a conventional microscope or share live images with your 
class using WiFi tablets or wireless laptops. View, capture, 

measure and edit images on iOS or Android devices with 
the free MotiConnect App, or through IP address using 

web browser and the camera’s on-board software.
 
You must charge batteries prior to using this instrument. 

See Recharging Batteries on page 10 (S).

See WiFi camera operation instructions located on page 
38.

To use and focus this microscope see General Instructions 
for Compound Microscopes located on pages 14 and 15.

LED bulb replacement:  See LED Bulb Replacement 
located on page 7 (D).

Battery replacement: See Replacing Batteries located on 
page 10 (T).

Содержание S7001B

Страница 1: ...Microscope Instruction Manual...

Страница 2: ...nses The image produced by the compound microscope is upside down and reversed Compounds are used for viewing standard 1 by 3 1mm thick transparent specimen slides with cover slips Prepared specimen s...

Страница 3: ...C Instructions 21 Model S71003D S71003E S71003F S71003G S71003H and S71003J Instructions 22 General Instructions for Stereo Microscopes 23 25 Model S71004 and S71004A Instructions 26 Model S71005 S710...

Страница 4: ...Disc located below stage with holes of various apertures designed to help achieve optimum resolution of the objective lens Larger apertures used for higher magnifications and smaller apertures used f...

Страница 5: ...st to the object being viewed forms first image of the specimen OIL IMMERSION LENS High power 100x objective lens which requires a medium of immersion oil between the lens and the slide RESOLVING POWE...

Страница 6: ...cleaning will be submitted for cleaning III MECHANICAL MAINTENANCE A Rack Stop compound microscopes only Rack stop screw has been pre adjusted at the factory and should not require re adjustment Howev...

Страница 7: ...and replace locking screw C Replacing Incandescent Lamps on Middle High School Models S71001 and S71001A a Make sure that the illuminator field lens housing and lamp are cool before attempting to remo...

Страница 8: ...3B S71003C S71003D S71003E S71003F S71003G S71003H S71003J S71012 and S71012A a Carefully lay instrument on its side taking care to avoid damage to specimen slide holder located on top of mechanical s...

Страница 9: ...l it slips into position d Replace stage plate and secure with locking setscrew M Replacing the Incidental Top Illuminator Halogen Bulb on Digital Stereo Model S71012B a Remove front lens housing from...

Страница 10: ...il slot engages depress fuse cap and rotate turn to lock into fuse holder S Recharging Batteries on Cordless Models a Must use the supplied Automatic Switching Recharger when charging batteries b It i...

Страница 11: ...Locate battery compartment cover on bottom of illuminator Carefully remove the two screws securing battery cover to bottom of illuminator Slide cover back to expose batteries d Remove all 3 batteries...

Страница 12: ...Incorrect bulb voltage or lamp base used Replace with specified lamp Light flickers Lamp not properly inserted into socket Loose connection at AC outlet Electrical short Properly insert lamp Have qual...

Страница 13: ...slide dirty Spots on field of view Eyepiece or condenser lens dirty No immersion oil used on 100X objective lens Bubbles air in immersion oils Adjust disc diaphragm or iris diaphragm Adjust light int...

Страница 14: ...entered over lens in center condenser lens located in middle of stage B Mechanical Stage Models Swing movable finger on holder outward place specimen slide cover slip facing up on top of slide against...

Страница 15: ...ything when looking through the microscope A slight adjustment of focus and diaphragm openings will be needed to sharpen the image when changing objective magnification A Models supplied with 100x oil...

Страница 16: ...hat you will want adequate ambient light in order to illuminate your opaque specimen from the top If the specimen is small and falls through the aperture at center of stage place a small piece of pape...

Страница 17: ...ble on model S71000D Battery replacement for models S71000D See Replacing Batteries located on page 10 U MODEL S71000E and S71000F These two models have a special inclined eyepiece tube so that it is...

Страница 18: ...1D Replacing batteries see page 10 T Fuse replacement on models S71001 and S71000A See Replacing Fuses located on page 10 R MODEL S71001E S71001F AND S71001G Dual viewing head permits simultaneous vie...

Страница 19: ...ated on page 10 T S71002A AND S71002B STUDENT MODEL These models feature coarse and fine focusing mechanisms full sized DIN 4X 10X and 40X objectives widefield 10X eyepiece and a 0 65 condenser with i...

Страница 20: ...Fuse replacement on models S71002C and S71002E See Replacing Fuses located on pages 10 R S71002G HIGH SCHOOL AND COLLEGE This model features a Seidentopf binocular head coaxial coarse and fine focusin...

Страница 21: ...e To operate this series of microscopes see General Instructions for use of a Compound Microscope located on pages 14 and 15 Lamp replacement For Halogen lamp replacement see page 8 H Fuse replacement...

Страница 22: ...of objective lenses After properly focusing instrument adjust ocular tube diopters to compensate for the difference between left and right eyes Lamp replacement For Halogen lamp replacement see page 8...

Страница 23: ...ion Select illumination that meets the requirements of specimen and turn ON illuminator A There are three rocker type light controls located on microscope base a MAIN Turns power ON and OFF b I Turns...

Страница 24: ...e for this difference and assure that you will see one corrected image when looking through microscope a Observe that the knurled diopter ring on the left eyepiece tube can be rotated to move the eyep...

Страница 25: ...knob to the lowest magnification by aligning the number 1 located on the objective turret for model S71010 or by aligning the number 1 located on zoom control knobs to the index dot on the viewing hea...

Страница 26: ...bottom lamp located on page 9 L Fuse replacement See Replacing Fuses located on pages 10 R MODEL S71004A Features an adjustable post mounted viewing head with inclined eyepiece tubes for ease in viewi...

Страница 27: ...harge batteries prior to using this instrument See Recharging Batteries page 10 Q To use and focus this microscope see General Instructions for Stereo Microscopes located on pages 23 25 Models S71005...

Страница 28: ...age 9 K Transmitted bottom lamp replacement See Replacing the transmitted bottom lamp located on page 9 L Fuse replacement See Replacing Fuses located on page 10 R MODELS 71007 and S71007A Features in...

Страница 29: ...Fuses located on page 10 R MODEL S71008A Fixed arm with inclined viewing head Illuminated by a variable halogen incidental top and a fluorescent transmitted bottom lighting system Change magnificatio...

Страница 30: ...located on page 9 O Fuse replacement See Replacing Fuses located on page 10 R MODEL S71010 Fixed arm with inclined viewing head Illuminated by a variable halogen incidental top and a fluorescent trans...

Страница 31: ...tic Images software allowing the user to view capture label manipulate and measure images In order to achieve optimum results it is important that you carefully read the instructions for software loca...

Страница 32: ...outlet and converter automatically switches to proper output voltage Refer to Quick Start Guide located on your CD for installation of the software and operation of the camera The three position slid...

Страница 33: ...matically switches to proper output voltage Refer to Quick Start Guide located on your CD for Installation of the software and operation of the camera The three position sliding rod located on the sid...

Страница 34: ...plug into 12VDC switching power converter then insert plug on other end of converter into power jack on back of microscope base Plug AC cord into either a 120v or 240v outlet and converter automatical...

Страница 35: ...ble magnification from10X to 40X Illumination Insert power plug into 12VDC switching power converter then insert plug on other end of converter into power jack on back of microscope base Plug AC cord...

Страница 36: ...cluded with microscope Instructions on connecting to your wireless enabled laptop or desktop are covered in the Wifi Camera Operation section of this manual MODEL S71014 This WiFi digital microscope t...

Страница 37: ...images on iOS or Android devices with the free MotiConnect App or through IP address using web browser and the camera s on board software You must charge batteries prior to using this instrument See...

Страница 38: ...warranty b Once the connection has been established you can begin using the WiFi camera through the MotiConnect App automatic or through your web browser using the following IP address http 192 168 1...

Страница 39: ...nician s discretion Other than set forth above Fisher Science Education hereby disclaims all warranties expressed or implied of fitness for a particular purpose Defective products covered by the warra...

Страница 40: ...tomer service call 1 800 955 1177 To fax an order use 1 800 955 0740 To order online www fisheredu com 2013 Thermo Fisher Scientific Inc All rights reserved Trademarks used are owned as indicated at w...

Отзывы: