![Fisher-Rosemount 3095FT Скачать руководство пользователя страница 126](http://html1.mh-extra.com/html/fisher-rosemount/3095ft/3095ft_manual_2300591126.webp)
Rosemount Model 3095FT HART Flow Transmitter
7-4
Process Temperature (RTD)
Specification for process temperature is for the transmitter portion
only. Sensor errors caused by the RTD are not included. The
transmitter is compatible with any PT100 RTD conforming to IEC
751 Class B, which has a nominal resistance of 100 ohms at 0 °C and
∝
= 0.00385. Examples of compatible RTDs include the Rosemount
Series 68 and 78 RTD Temperature Sensors.
Range
–40 to 185 °F (–40 to 85 °C).
Accuracy (including Linearity, Hysteresis, Repeatability)
±1.0 °F (0.56 °C).
Ambient Temperature Effects per 50 °F (28 °C)
±0.72 °F (0.40 °C).
Stability
±1.0 °F (0.56 °C) for 12 months.
MODEL 3095FT
PHYSICAL
SPECIFICATIONS
Electrical Connections
½–14 NPT, CM 20, PG-13.5.
Process Connections
Transmitter: ¼–18 NPT on 2
1
/
8
-in. centers.
RTD: RTD dependent (see ordering information).
Process Wetted Parts
Isolating Diaphragms
316L SST or Hastelloy C-276.
®
Drain/Vent Valves
316 SST or Hastelloy C.
®
Flanges
Plated carbon steel, 316 SST, or Hastelloy C.
Wetted O-rings
Glass-Filled TFE.
Non-Wetted Parts
Electronics Housing
Low copper aluminum.
Bolts
Plated carbon steel per ASTM A449, Grade 5;
or austenitic 316 SST.
Fill Fluid
Silicone oil.
Paint
Polyurethane.
O-rings
Buna-N.
Weight
Component
Weight in lb (kg)
Model 3095FT Transmitter
LCD Meter
SST Mounting Bracket
12 ft (3.66 m) RTD Cable
24 ft (7.32 m) RTD Cable
6.0 (2.7)
0.5 (0.2)
1.0 (0.4)
0.5 (0.2)
2.2 (1.0)
Содержание 3095FT
Страница 1: ...R Model 3095FT HART Flow Transmitter 00809 0100 4015 English Rev B1...
Страница 2: ......
Страница 4: ......
Страница 45: ...3 13 Options and Accessories 3051 1006D03A FIGURE 3 5 Recharging the Battery 15 to 36 V dc 500 mA minimum...
Страница 98: ...Rosemount Model 3095FT HART Flow Transmitter 4 50...
Страница 117: ...5 19 Troubleshooting and Maintenance FIGURE 5 2 Replacing the Remote Power Supply Battery...
Страница 118: ...Rosemount Model 3095FT HART Flow Transmitter 5 20...
Страница 146: ...Rosemount Model 3095FT HART Flow Transmitter 7 24...
Страница 147: ...Appendix A 1 A Model 3095 Explosion Proof Installation Drawing Factory Mutual Drawing Number 03095 1025 Rev G...
Страница 148: ...Rosemount Model 3095FT HART Flow Transmitter A 2...
Страница 149: ...A 3 Appendix A...
Страница 150: ...Rosemount Model 3095FT HART Flow Transmitter A 4...
Страница 151: ...Appendix B 1 B Index of I S F M for 3095 and Index of I S F M for 03095 5000 Drawing Numbers 03095 1020 and 03095 5000...
Страница 152: ...Rosemount Model 3095FT HART Flow Transmitter B 2...
Страница 153: ...B 3 Appendix B...
Страница 154: ...Rosemount Model 3095FT HART Flow Transmitter B 4...
Страница 155: ...B 5 Appendix B...
Страница 156: ...Rosemount Model 3095FT HART Flow Transmitter B 6...
Страница 157: ...B 7 Appendix B...
Страница 158: ...Rosemount Model 3095FT HART Flow Transmitter B 8...
Страница 159: ...Appendix C 1 C Model 3095 Explosion Proof Installation Drawing CSA Drawing Number 03095 1024 Rev D...
Страница 160: ...Rosemount Model 3095FT HART Flow Transmitter C 2...
Страница 161: ...C 3 Appendix C...
Страница 162: ...Rosemount Model 3095FT HART Flow Transmitter C 4...
Страница 163: ...Appendix D 1 D Index of I S CSA for 3095 and Index of I S CDA for 03095 5000 Drawing Numbers 03095 1021 and 03095 1031...
Страница 164: ...Rosemount Model 3095FT HART Flow Transmitter D 2...
Страница 165: ...D 3 Appendix D...
Страница 166: ...Rosemount Model 3095FT HART Flow Transmitter D 4...
Страница 167: ...D 5 Appendix D...
Страница 168: ...Rosemount Model 3095FT HART Flow Transmitter D 6...
Страница 183: ......