List of Task Commands
Task programming
FISCHERSCOPE
®
X-RAY
81
p62V2Sets variable 2 = V3
p63V3Sets variable 3 = V3
p71L V1 = V1 -1, if V1 > 0 GOTO label L
p72L V2 = V2 -1, if V2 > 0 GOTO label L
p73L V3 = V3 -1, if V3 > 0 GOTO label L
Accessories
f42 S Creates a text box on the screen (Text = S)
Returns TRUE or FALSE (Button [OK] or [Cancel])
f50 - Starts a reference measurement
f64 N Sets a delay (pause) of N sec.
f90 - Closes a window if not the main window
(If no window can be closed then return code = ERROR)
f91 - Reset X-RAY, new PC-Communication, new HV
control etc.
p98- Singlestep mode: Task is carried out by single steps
following F4 key print
p99- Singlestep mode (p98) OFF
.. depending on the last readings (tolerance limits) ...
p80S,LS symbolizes a logical relation acc. to the table given
below. In case of TRUE (the respective relation is
fulfilled) the task steps to label L.
... depending on the last measurement data ... (specification
limit)
p80S,LS symbolizes a logical correlation according to the
table below.
In case of TRUE (the respective relationship is fulfilled), the task
will jump to label L.
S Meaning
(“i” represents the number of the measurement data column)
> KiReading no. i (i=1,2,3) > USL
< KiReading no. i (i=1,2,3) < LSL
> KXminimum 1 reading is > USL
< KXminimum 1 reading is < LSL
?Ki Reading no. i (i=1,2,3) is not measurable
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...