Variables Available for the final Evaluation
Setting Up the Print Form Using PDM
FISCHERSCOPE
®
X-RAY
65
@ANBOrder number of
the last block of the selected blocks
or the information that has been entered into the first entry
field of the dialog window “Order No./ Operator”.
@BM1Lot number of
the last block of the selected blocks
or the information that has been entered into the second entry
field of the dialog window “Order No./ Operator”.
@EDBNumber of
the last block of the selected blocks
.
@EDTDate of
the last block of the selected blocks
.
@ETMTime of
the last block of the selected blocks
.
@HG1Histogram of all single readings of the selected blocks
in the current product.
See also @WN1.
Regarding the numerals
1; 2; 3 ... after certain
variables:
A measurement consists
of one or more single
readings (“channels”) that
are identified by consecu-
tive numbers (1,2,3).
Example “Dual coating
measurement”
Top coating thickness = 1
Intermediate coating
thickness = 2
@OE1Upper control limit UCL(d) of the mean values of an
SPC chart.
@OG1Upper specification limit USL (upper tolerance) for the
presentation and computation of the upper limit
violations in the selected blocks of the current product.
@OR1Upper control limit UCL(s) or UCL(R) of the standard
deviation s or the range R of an SPC chart.
@P<1Percentage of single readings in the selected blocks of the
current product that fall below the lower specification limit
LSL.
@P>1Percentage of single readings in the selected blocks of the
current product that exceed the upper specification limit
USL.
@PRFOperator of the last block of the selected blocks
or the information entered into the third entry field of the
dialog window “Order No./ Operator”.
@RG1Range = Difference between the largest (maximum) and
smallest (minimum) single readings of the selected blocks
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...