FISCHERSCOPE
®
X-RAY
121
Calibration
Type
With multi-coating systems, it may be the case that existing
interim coatings act as base material of a calibration stan-
dard or that coatings are not present. This setting is made
here.
7. Once you have entered all values of a calibration standard, click
OK
.
8. To enter additional calibration standards, repeat steps 3. to 7. for every calibration
standard. Before you enter an additional calibration standard, make sure that no line
is selected in the window
Input Calibr. Std. Set
. Otherwise the values of the selected
standard will be changed.
9. Once you have entered all values of a calibration standard, click
OK
in the window
Input Calibr. Std. Set
.
The calibration standard set is now entered and stored.
10. To check how many measurements will be performed
for each calibration standard, click
Info
.
The window
Modify product
appears.
11. Click the tab
Applica-
tion
.
12. Check the following
values:
T_cal (sec)
: Time for each individual measurement on the calibration standard
N_cal
: Number of individual measurements on each calibration standard
Typically, the default setting applies but you do have the option of changing the values.
The following applies to the measuring time
T_cal
:
The longer the measurement, the better the repeatability precision. Do not select
a measuring time that is too short; 30 or 60 seconds is a typical value.
The following rules apply to the number of individual measurements
N_cal
:
Pure elements, base materials or scatter spectra need to be measured only once.
Calibration standards must be measured several times on each standard, at least
three times but typically five times. Perform the measurements at different loca-
tions on the calibration standard.
13. To start the calibration, select
Calibrate > Calibrate...
.
WinFTM will prompt you to position the calibration stan-
dards in succession and to measure them.
Example: The calibration standard Cr/Ni/Cu with a
coating thickness of 2.58
μ
m shall be measured now.
NOTE: To read the full name of the calibration standard,
click in the row and use the arrow keys to scroll back
and forth.
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...