120
FISCHERSCOPE
®
X-RAY
Calibration
Procedure
1. In the WinFTM main window select
Calibrate > Calibration Standard Set...
An empty window will
appear, if the measuring
application has not yet
been calibrated.
If the measuring applica-
tion has already been
calibrated, check the
values in the window. If
the values are plausible,
click
OK
and skip the
2. Enter a name for the cali-
bration standard set in
the field
Label
(e.g., Cr/
Ni/Cu Ser. No. 13917).
3. Click
Input/Overwrite.
4. Tick the option
Element-Corr.
If this option is used, the pure element spectra are determined during calibration.
The window
Standard No. 1
appears.
5. Enter the name of the
standard
(e. g., AAMGW).
6. Enter the values of the
calibration standard.
Click the field three times
in order to enter a value.
The fields have the
following meaning:
Value
Nominal value of the calibration standard (in the example
coating thickness d1 (Zn) in
μ
m coating thickness d2 (Ni) in
μ
m, Cu concentration in %)
u +/-
If an uncertainty for the reference value is stated on the
certificate for the calibration standard, enter this value here.
If you do not enter a value, then the following values will be
used: for coating thicknesses, 1% of the coating thickness;
for concentrations, 1% absolute.
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...