mq Value
User Interface of the WinFTM Software
5.6.1
Upper Limit of the mq Value
The factory default for the upper limit of the mq value is set to 5. You do have the option
to change this upper limit.
When the upper limit of the mq value is exceeded during a measurement, the measure-
ment will be terminated and the error message
ERROR: Spectrum not suitable!
will be
issued. The related measurement value will not be added to the result list.
To change to upper limit of the mq value
1. In the WinFTM main menu, click
.
The window
Modify product
appears.
2. Select the tab
Application
.
3. In the field
mq_Max
, enter the new value for the upper limit.
If a value of 0 is shown in the field
mq_Max
, WinFTM will automatically set the mq value
to 5.
5.6.2
Possible reasons for mq values that are too high
The specimen is not correctly defined, e.g.:
wrong measurement application or wrong Def.MA
the expected substrate material does not correspond to the actual composition
interim coatings are not taken into account
Other reasons that may have influenced the measured spectrum in an unforeseeable
manner are possible as well, e.g.:
Shading of the radiation on its path from the sample to the detector through protruding
parts of the specimen
poor sample positioning
complex geometry of the specimen, curvature on the specimen too strong
measurement area too small
out of focus
Содержание FISCHERSCOPE X-RAY 5000 Series
Страница 18: ...18 FISCHERSCOPE X RAY Switching the Instrument on and off Communication between Instrument and WinFTM...
Страница 22: ...22 FISCHERSCOPE X RAY Performing Manual Measurements Deleting Measurement Readings...
Страница 36: ...36 FISCHERSCOPE X RAY User Interface of the WinFTM Software The Spectrum Window...
Страница 40: ...40 FISCHERSCOPE X RAY WinFTM File Structure Product...
Страница 112: ...Measurement device monitoring for the Fischerscope X RAY Long term monitoring 112 FISCHERSCOPE X RAY...
Страница 118: ...118 FISCHERSCOPE X RAY Def MA Display the Measurement Mode...
Страница 124: ...124 FISCHERSCOPE X RAY Calibration...
Страница 142: ...142 FISCHERSCOPE X RAY Addendum Periodic Table of the Elements with X Ray Properties...
Страница 156: ...156 FISCHERSCOPE X RAY Addendum Assignment of the Electrical Connections...
Страница 182: ...182 WinFTM WinFTM SUPER For the Experienced X RAY User Comparative Overview with without WinFTM PDM...
Страница 183: ...WinFTM 183...