15
F9
Weight Unit Enable / Disable
kg
(
** On
/Off)
g
(On/
** Off
)
lb
(On/
** Off
)
F10
Filter & Conversion Speed
1
2
3
4
** 5
6
7
8
9
Press
[FUNC/SET]
or
[MR/UNIT]
to select: -
1 (strongest filter) for bad working environment where vibration, wind flow… etc. affect stable reading,
5 for normal environment,
9 (least filter) for very good working environment where wind and vibration have no effect to stable reading.
Then press
[ZERO]
to quit or
[PRINT/M+]
to save and continue conversion speed setting.
4 parameters are available. Press
[FUNC/SET]
or
[MR/UNIT]
to select AD conversion speed: -
** 15
times per second.
30 times per second. Recommended maximum resolution = 15,000 division
60 times per second. Recommended maximum resolution = 7,500 division
120 times per second. Recommended maximum resolution = 3,000 division
Note
: - After AD conversion speed is changed, instrument will automatic restart.
F11
Auxiliary Function
Cnt (Counting) On /
** Off
AtM (Action Tare Memory) On /
** Off
Ani (Animal) On /
** Off
PCd (Quick Access to Customer & Product Code Setting) On /
** Off
OP (Operator code Input through keyboard during operation) On/**
Off
F12
Auto Tare Function
** OFF
ON
Contin
Notes: -
Off = Auto Tare Function disable.
On = Only the first table weight applied will be tare off. Minimum tare load ≥ 2d.
Contin = All stable weight applied will be tare off. Minimum tare load ≥ 10d.
If Contin is set, select also d.t. (delay time, 0.0 ~ 9.9 second.
Default = 0.5 second
). Delay time is the time duration from
when
a stable weight is detected until it is automatically tare off.
Enter the preferred d.t. value through numeric keys and then press
[PRINT/M+]
to save.
F13
Repetitive Tare Function
OFF
** ON
If F12 is set = Contin, Repetitive Tare setting “Off” will be surpassed.