5.4 Internal Function Table
Refer to the below tables for internal function number, parameter and setting notes.
No.
Description
Parameters / Note
Default = **
F1
Internal Analogue to Digital
(AD) Value
Press
[PRINT/M+]
to set offset value to zero when unloaded. Then add load on the platform to observe
the span value of load applied.
When ADC is more than 1 million, Zero Indicator will appear. Actual ADC is = 1 million
plus the ADC value being displayed.
When ADC is more than 2 million, Net Indicator will appear. Actual ADC is = 2 million
plus the ADC value being displayed.
When ADC is more than 3 million, kg Indicator will appear. Actual ADC is = 3 million plus
the ADC value being displayed.
Press
[ZERO]
to quit to F1.
F2
All Segment & Battery Voltage
Check
All display segments and indicators will be lit on. Check any segments or indicators are missing.
To check battery voltage, press
[Check]
.
F3
Capacity, Division & Default
Weight Unit
Display basic metrology characteristics (capacity, division, and weight unit). Value displayed = Max +
1e
F4
Date Format & Date
DD/MM/YY
** YY/MM/DD
MM/DD/YY
•
Press
[PRINT/M+]
to check current date format.
•
To check/change date value, press
Print/M+]
again then enter date value and then press
Print/M+]
to confirm.
F5
Time
HH/MM/SS
To change time, press
[PRINT/M+]
, then enter a new value and press
[PRINT/M+]
.
F6
System Initialization
(Set F7~F32) to Default)
** NO
YES
If YES is selected, press
[1]
when “
SURE?
” is displayed. Indicator shows Done when initialization is completed.
F7
Auto Power Off Time (Minute)
OFF
1
3
** 5
10
20
Auto power off function will be disabled when an energized power adaptor is plugged in.
F8
Brightness
1
2
3
** 4
5
1 = dimmest; 5 = brightest
Press
[FUNC/SET]
or
[MR/UNIT]
to change setting and then press
[PRINT/M+].