Fusion splicer owns the capacity to finish routine maintenance, this chapter will
expound how to use maintenance menu.
④
Please use clean cloth that contain alcohol or soft lint to clean the new electrode
and then install it to the Fusion Splicer
。
(i) Put them under the electrode cover.
(ii) Take out the electrode.(Put the electrode to the electrode cover).
①
Select “Maintenance” in main menu and then press to enter into the
maintenance menu.
②
Select the function to execute.
The e lectrode w ill abrasion due t o long-time use, s o you m ust d o regular
cleanness according to the aggregation extent of the oxide. We suggest that the
electrode should be r eplaced a fter discharging one t housand t imes. W hen t he
electrode discharging time up to one thousand times, the message to remind you
to r eplace the electrode w ill b e appear. If you don’t replace t he e lectrode a fter
long-time use will result in splice loss increase and splice strength reduction.
Procedure to replace the electrode:
①
Implement“replace electrode”in“maintenance menu”option.
②
he screen will show the message to turn off the power, press until LED shut
down.
③
Remove the old electrode.
(i) Loosen the screw that fixed to the electrode lid.
(ii) Take out the electrode.(Put the electrode to the electrode cover).
Chapter 7 Maintenance Menu
7.1 Replace the electrode
Loosen the screw
Replace the old electrode
with the new electrode
Tighten the screw
Electrode cover
Note:
Please not pull the wire out when replace the electrode.
The torque should be 2kgf-cm when tight the screws.
Loosen the screw