10-5
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Figure 10-5Alarm event review window
What displayed in this window are follows:
Alarm Time
Alarm event
Alarm level
The waveform and parameter value at the alarm moment
Attention
When the alarm events are more than 200, the monitor will only store the latest 200 events
and the earlier ones will be overwritten.
The alarm log is not maintained when the alarm system is powered down and the time of
powering down is not captured in the alarm log, neither. All the physiological alarm
events will be cleard after the alarm system has experienced a total loss of power for a finite
duration.
10.5 Waveform Review
The monitor can save 120 minute parameter waveform for review in the waveform review window.
How to perform waveform review:
1.
Click [INFO RECA.] on-screen key;
2.
Select [WAVE RECALL] to enter the [WAVE RECALL] menu, as shown below: