III-1
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Appendix III System Alarm Message
(1) Physiological alarm message
Alarm Message
Default
alarm level
Selectable
Alarm Level
Causes
Solution
FHR1TOO HIGH
Medium
High, Medium,
Low
The measured parameter
value is higher than the
upper alarm limit or lower
than the lower alarm limit.
Check the patient’s
condition and check if the
alarm limit is suitable for
the patient type.
FHR1TOO LOW
Medium
High, Medium,
Low
FHR2 TOO HIGH
Medium
High, Medium,
Low
FHR2 TOO LOW
Medium
High, Medium,
Low
TOCO TOO HIGH Medium
High, Medium,
Low
TOCO TOO LOW
Medium
High, Medium,
Low
HR TOO HIGH
Medium
High, Medium
HR TOO LOW
Medium
High, Medium
ST1 TOO HIGH
Medium
High, Medium
ST1 TOO LOW
Medium
High, Medium
ST2 TOO HIGH
Medium
High, Medium
ST2 TOO LOW
Medium
High, Medium
PVC TOO HIGH
Medium
High, Medium
PVC TOO LOW
Medium
High, Medium
SPO2 TOO HIGH
Medium
High, Medium
High, Medium
SPO2 TOO LOW
Medium
PR TOO HIGH
Medium
High, Medium
PR TOO LOW
Medium
High, Medium
NS TOO HIGH
Medium
High, Medium
NS TOO LOW
Medium
High, Medium
ND TOO HIGH
Medium
High, Medium
ND TOO LOW
Medium
High, Medium
NM TOO HIGH
Medium
High, Medium
NM TOO LOW
Medium
High, Medium
RR TOO HIGH
Medium
High, Medium
RR TOO LOW
Medium
High, Medium
T1 TOO HIGH
Medium
High, Medium
T1 TOO LOW
Medium
High, Medium
RESP APNEA
High
High
The respiration signal is
so weak that the monitor
cannot perform respiration
Check
the
patient’s
condition and check the
connection of lead wires.
Содержание FM5000C
Страница 75: ...14 3 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure14 2 FHR Transducer ...
Страница 83: ...14 10 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual patient to avoid cross infection ...
Страница 97: ...15 13 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual baseline ...
Страница 102: ...16 1 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure 16 2 Positions of electrodes of 5 lead ECG cable ...
Страница 148: ...IV 2 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual PR alarm limit Lower limit 45 70 110 ...
Страница 155: ...Product Introduction Rear View Handle Paper tray Battery compartment Alarm Speaker Fetal Heart Speaker Nameplate ...