II-4
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
HR Alarm range
Adult
15~300bpm
Pediatric/Neonate
15~350bpm
Alarm error
±
1bpm
Alarm start time for
asystole
<10s
Alarm start time for
heart rate too low or
too high
<10s
Input dynamic range
Input signal amplitude
±5mV
Rate (RTI)
320mV/s
DC offset voltage
-300 ~ +300mV
Output signal change
±10%
Failure display (attenuation
prior to display)
Not reduced to below 50%
Input impedance
5M
System noise
(p-v
RTI)
<30µV
Multi-channel
crosstalk
<5%
Time
reference
selection and accuracy
Time reference
selection
Permanent display
25mm/s
Non-permanent display
12.5mm/s, 25mm/s, 50mm/s
Accuracy
±10%
Output display
Channel width
30mm
Aspect ratio
0.4s/mV
Input
signal
reproduction accuracy
Total system error
±20% or ±100
µ
V, whichever is larger
Frequency response
Sinusoidal input
0.67~40Hz
(attenuation-3dB)
Response to 20ms
(width) triangular wave
0~25Hz
attenuation
in
amplitude of wave peak
Response to 0.3 mV·s
shock in the shock
range
Offset (RTI)
≤0.1mV
Slope (RTI)
≤0.30mV/s
Electrode weighting
factor
≥±5%
Hysteresis effect of
15mm offset
≤0.5mm
Calibration voltage
1mV , error: ±5%
Common
mode
rejection
<1mV (p-v RTI)
Содержание FM5000C
Страница 75: ...14 3 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure14 2 FHR Transducer ...
Страница 83: ...14 10 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual patient to avoid cross infection ...
Страница 97: ...15 13 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual baseline ...
Страница 102: ...16 1 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure 16 2 Positions of electrodes of 5 lead ECG cable ...
Страница 148: ...IV 2 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual PR alarm limit Lower limit 45 70 110 ...
Страница 155: ...Product Introduction Rear View Handle Paper tray Battery compartment Alarm Speaker Fetal Heart Speaker Nameplate ...