8-8
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
8.5.4
Setting the Alarm Delay Time
The alarm system provides five kinds of alarm delay time: [DISABLED], [5SEC], [10SEC], [15SEC] and
[20SEC]. [DISABLED] indicates that the monitor will give alarms immediately once the measured
parameter value violated the alarm limit; [5SEC] indicates the monitor will give alarms only when the
measured parameter value exceeds the alarm limit for continual 5 seconds; and so forth. The alarm delay
function is not available on ECG parameter.
How to set the alarm delay time:
1. Access [MAIN MENU], select [MAINTAIN] and enter the password;
2. Select [ALARM SETUP]
3. Select [ALM DELAY TIME] to adjust the alarm delay time.
8.6 Pausing Alarms
The users can press the
key on the front panel or click [ALM PAUSE] on-screen key to pause the alarm
triggered and put the alarm system into alarm paused status.
1.
Stop all the indications of physiological alarms and no new physiological alarm will be triggered.
2.
The remaining pause time “ALM PAUSE xxS” will appear at the physiological alarm message area.
3.
[ALM PAUSE] on-screen key flashes in red.
4.
For technical alarm, the audio and visual alarm indicationwill be paused but alarm message still appears in
technical alarm message area. If new technical alarm is triggered during alarm pause, only the alarm
message of this technical alarm will be displayed without audio and visual alarm indication.
When the remaining alarm pause time ends, the alarm paused status will be automatically cancelled. You can
also cancel the alarm paused status by pressing the
key on the keyboard or clicking [ALM PAUSE]
on-screen key.
8.6.1
Setting the Maximum Alarm Pause Time
1.
Access [MAIN MENU], select [MAINTAIN] and enter the password;
2.
Select [ALM SETUP]
3.
Select [MAX ALM PAUSE TIME] to adjust the maximum alarm pause time. Available options are 2
minutes to 10 minutes.
8.6.2
Setting the Alarm Pause Time
1.
Access [MAIN MENU]
Содержание FM5000C
Страница 75: ...14 3 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure14 2 FHR Transducer ...
Страница 83: ...14 10 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual patient to avoid cross infection ...
Страница 97: ...15 13 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual baseline ...
Страница 102: ...16 1 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure 16 2 Positions of electrodes of 5 lead ECG cable ...
Страница 148: ...IV 2 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual PR alarm limit Lower limit 45 70 110 ...
Страница 155: ...Product Introduction Rear View Handle Paper tray Battery compartment Alarm Speaker Fetal Heart Speaker Nameplate ...