9-3
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
(a) Disconnect the monitor from the patient and suspend all monitoring and measuring procedures.
(b) The optimized battery should be kept in the battery compartment of the unit.
(c) When charging the battery, at least six hours of charging should be ensured until it is fully charged.
(d) When you disconnect the AC power supply, the monitor is powered with the battery until the
battery runs out and the monitor automatically shuts off.
(e) This completes the battery optimization process.
(2) Check of Battery Performance
The service life of battery is changeable along with its storage, working environment charge cycles and
service time. Even though battery is out of service its performance will gradually deteriorate.
Procedure for checking the battery is as follows:
(a) Confirmwhether or not the battery is damage. When the battery shows the symbol “
”, it
indicates the battery is damaged or no battery.
(b) Check whether the battery can be normally charged when the battery is connected to alternating
current;
(c) Disconnect the monitor from the patient and suspend all monitoring and measuring procedures.
(d) When charging the battery, at least six hours of charging should be ensured until it is fully charged.
(e) Disconnect the AC power supply, power on the monitor with the battery until it is fully discharged
and the monitor shuts off automatically Record the start and stop time.
(f) The duration of battery discharge will reflect the battery performance.
(g) Once the discharge duration is down to 50% of the original time, it requires changing the battery.
Attention
In order to extend the service life of the battery, it is recommended to charge it every three
months after a long dormant period so as to prevent overdischarge.
Battery power supply loss depends on the configuration and operation of the monitor; for
example, the unit will have a big loss of battery power if it is used to measure NIBP parameter
often.
9.6 Battery Recovery
If the battery shows apparent damage or is at an energy exhaustion condition, it should be exchanged
immediately, and the used battery should be recovered and properly disposed of in accordance with relevant
laws or rules and regulations for hospitals.
Warning
Do not remove the battery or make it short-circuiting or put it into fire; otherwise, it would
cause battery on fire, explosion, harmful gas leakage or other dangers.
Содержание FM5000C
Страница 75: ...14 3 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure14 2 FHR Transducer ...
Страница 83: ...14 10 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual patient to avoid cross infection ...
Страница 97: ...15 13 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual baseline ...
Страница 102: ...16 1 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure 16 2 Positions of electrodes of 5 lead ECG cable ...
Страница 148: ...IV 2 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual PR alarm limit Lower limit 45 70 110 ...
Страница 155: ...Product Introduction Rear View Handle Paper tray Battery compartment Alarm Speaker Fetal Heart Speaker Nameplate ...