9.3 Power Management
The power management function is designed to save power and reduce emission for the sake of
environmental protection.
Click battery icon to enter [POWER MANAGE] menu and set the following items:
[TIMER]: click to select time interval of from 1 minute to 30 minutes, within which if no operation
command is received, the monitor will trigger a command set in [TIMER CALL].
[TIMER CALL]: set the command that the monitor will trigger if no operation command is
received within the time interval set in [TIMER], including [NONE], [HIDE BTTM BTTN](hide
bottonshortcut keys), [LOCK SCREEN], [STANDBY] or [SHUTDOWN].
[RECOVER FROM]: select [TOUCH SCREEN], [KEY] or [ANY OPERAT](any operation) as the
method of making the monitor return to the main screen after a command set in [TIMER CALL] is
triggered.
9.4 Installing Battery
Procedure of changing or installing the battery:
(1) Turn off the monitor, and disconnect the power cord and other connection lines.
(2) Place the monitor with the back up.
(3) Unscrew the battery cover.
(4) Take out the used battery and put the new one into the battery holder making sure the positive and
negative poles match what is indicated on its shell.
(5) Replace the battery and screw it on, and turn the monitor upright.
Warning
Use only the supplier’s designated battery.
Don’t dismantle the battery while the monitor is turned on.
The installation and replacement of the battery should be performed by trained and qualified
personnel.
9.5 Optimization and check of battery performance
(1) Optimization of battery performance
When the battery is used for the first time at least two complete cycles of optimization of the battery should be
carried out. A complete optimization cycle should be:uninterrupted charging battery until the power is full,
followed by use until the battery is fully discharged and monitor is automatically shut off.
This will ensure the battery is in optimization process:
Содержание FM5000C
Страница 75: ...14 3 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure14 2 FHR Transducer ...
Страница 83: ...14 10 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual patient to avoid cross infection ...
Страница 97: ...15 13 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual baseline ...
Страница 102: ...16 1 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual Figure 16 2 Positions of electrodes of 5 lead ECG cable ...
Страница 148: ...IV 2 FM5000C FM5000PLUS Maternal Fetal Monitor User Manual PR alarm limit Lower limit 45 70 110 ...